Catalog Number D140401
Device Problems Material Separation (1562); Communication or Transmission Problem (2896); Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a lassostarnav,10p,15mm dia loop and the catheter fell apart.It was reported that the lassostarnav catheter fell apart and is currently stuck inside the cryo balloon.The catheter was replaced.There was no patient consequences.
 
Manufacturer Narrative
Device investigation details: additional information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device 30981727l number, and no internal action related to the complaint was found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4) on 3-jan-2024, iit was noticed the following information was inadvertently omitted from the 3500a initial mdr: "resistance was felt during retraction of the catheter.Damage resulted in wire exposure and the device was not used with the heliostar catheter.Also, the device was also not being recognized by the carto 3 system.They tried 20-b and also replacing the dongle but the issue is still there." as such, h6.Medical device problem code of "communication or transmission problem (a13) has been added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASSOSTARNAV,10P,15MM DIA LOOP
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18423292
MDR Text Key331932281
Report Number2029046-2024-00001
Device Sequence Number1
Product Code DRF
UDI-Device Identifier10846835020335
UDI-Public10846835020335
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211219
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/13/2024
Device Catalogue NumberD140401
Device Lot Number30981727L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CRYO BALLOON.
-
-