Catalog Number 2426-0500
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd alaris pump module smartsite infusion set was missing component.The following information was received by the initial reporter with the following verbatim.Aah internal id number (b)(4).Date event occurred: 11/29/2023.Aah facility name: (b)(6) medical center.Product name(s): bd alaris pump infusion set (primary set).Manufacturer name: bd.Manufacturer number(s): (b)(4).Lot number(s): 1023103051.Quantity impacted: 1.Description of event/concern: primary iv tubing missing entire male luer.
 
Event Description
No additional information was provided.
 
Manufacturer Narrative
Dhr.No product or photo was returned by the customer.The customer complaint of entire male luer missing could not be verified due to the product not being returned for failure investigation.Device history record review for model 2426-0500 lot number 23103051 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 11oct2023.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause of this failure could not be identified without a failure investigation.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS PUMP MODULE SMARTSITE INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18423418
MDR Text Key331931730
Report Number9616066-2023-02450
Device Sequence Number1
Product Code FPA
UDI-Device Identifier37613203021006
UDI-Public(01)37613203021006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2426-0500
Device Lot Number23103051
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-