• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134702
Device Problems Material Puncture/Hole (1504); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an unspecified ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post-procedure the bwi product analysis lab identified a cut on the pebax.During the procedure, the catheter broke 10cm from the handle.No further details were provided regarding the rest of the procedure.No patient consequences were reported.
 
Manufacturer Narrative
E1 initial reporter phone: (b)(6) the product investigation was completed.Device evaluation details: the product was returned to biosense webster (bwi) for evaluation.Bwi conducted a visual inspection and deflection evaluation of the returned device.Visual analysis of the returned sample revealed a cut on pebax with reddish-brown material inside and internal parts exposed; also, the shaft was bent, however, these conditions could be related to the handling since in the process there are control inspection points to avoid this kind of issue; however, this could not be conclusively determined.Then deflection testing was performed and observed that the tip deflects properly.A manufacturing record evaluation was performed for the finished device 30787529l number, and no internal action was found during the review.The customer complaint was confirmed due bent shaft.However, the reddish brown material (possibly blood) is unrelated to the customer's initial complaint.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18428414
MDR Text Key331933337
Report Number2029046-2024-00019
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009781
UDI-Public10846835009781
Combination Product (y/n)N
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134702
Device Lot Number30787529L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2023
Date Manufacturer Received12/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-