MAUDE Adverse Event Report: ZIMMER BIOMET, INC. DISPOSABLE MIXING BOWL AND SPATULA; HIPS, INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. DISPOSABLE MIXING BOWL AND SPATULA; HIPS, INSTRUMENTS Back to Search Results
Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).G2: foreign: japan customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during incoming inspection that a team member found a hair like substance.There was no patient involvement.It was reported that no further information is available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Visual evaluation of the returned product found a hair-like fiber in the sealed sterile packaging that is not acceptable.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.This complaint has been confirmed by evaluation of the returned product.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE MIXING BOWL AND SPATULA
Type of Device
HIPS, INSTRUMENTS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18429301
MDR Text Key331886307
Report Number0001822565-2024-00010
Device Sequence Number1
Product Code JDZ
UDI-Device Identifier00889024376564
UDI-Public(01)00889024376564(17)280930(10)66330621
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00504901100
Device Lot Number66330621
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-