MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number PM3562
Device Problems Capturing Problem (2891); Communication or Transmission Problem (2896); Wireless Communication Problem (3283)
Patient Problem Insufficient Information (4580)
Event Type  Injury  
Event Description
It was reported that telemetry lockout with difficulty in programming and inconsistent capture threshold was observed on the device.The device was explanted and replaced.
 
Manufacturer Narrative
The reported event of capture issue, telemetry lockout, and inability to program were not confirmed.The device was received with normal output and telemetry communication.Analysis performed including capture, programming, and rf and inductive communication tests did not identify any functional issues.Longevity assessment found the device was operating at near beginning-of-life voltage with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18431576
MDR Text Key331790673
Report Number2017865-2024-00146
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM3562
Device Lot NumberA000146517
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET LV LEAD; TENDRIL RA LEAD; TENDRIL RV LEAD
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-