TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801763 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heart lung machine (hlm) monitor displayed a 'system failure' message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The field service representative (fsr) performed mechanical, electrical, and mechanical testing and was not able to duplicate the reported complaint.Per the data log review by the manufacturer's technical support specialist, there was an indication that a corrupt event occurred on 07dec2023 which resulted in a shutdown.The unit operated to the manufacturer's specifications.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|