AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-Q290
Device Problems Leak/Splash (1354); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The customer's complaint was confirmed due to an immersed plug unit.Additionally the following findings were discovered; the water tightness was lost due to a cut on universal cord, the plug unit had corrosion due to water leakage, and the insertion tube was scratched.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the bronchovideoscope had an e315 (non-compatible endoscope) error reported.The issue was found during reprocessing for a diagnostic procedure.There were no reports of patient harm.The procedure was completed using a similar device.
 
Manufacturer Narrative
This supplemental report is being submitted to provide information that was inadvertently left out of the initial medwatch report and to provide the results of the legal manufacturer's final investigation.A review of the device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation the root cause could not be identified; however, it is likely that water invaded scope connector causing a short circuit at the circuit board.The event can be prevented by following the instructions for use which state: "important information ¿ please read before use - precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.- precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly, or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system - inspection of the endoscopic image confirm that the wli and nbi endoscopic images (except bf-mp290f) are normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2, ¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18432139
MDR Text Key332384944
Report Number9610595-2024-00084
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-Q290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Was the Report Sent to FDA? No
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-290 EVIS LUCERA ELITE VIDEO SYSTEM CENTER.
-
-