Model Number G158 |
Device Problems
High impedance (1291); Under-Sensing (1661)
|
Patient Problem
Chest Pain (1776)
|
Event Date 12/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was undersensing fine atrial fibrillation (af) and the non-boston scientific right atrial (ra) lead exhibited high, out-of-range pace impedance measurements of greater than 2000 ohms.The patient has also reported feeling vibrations from their chest.The device and non-boston scientific leads remain in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|
Event Description
|
It was reported that the device was undersensing fine atrial fibrillation (af) and the non-boston scientific right atrial (ra) lead exhibited high, out-of-range pace impedance measurements of greater than 2000 ohms.The patient has also reported feeling vibrations from their chest.The device and non-boston scientific leads remain in service.No adverse patient effects were reported.
|
|