Catalog Number 8065977763
Device Problem Scratched Material (3020)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
A healthcare professional had reported that, during an intraocular lens (iol) implant procedure, a common issue arose with the cartridges.There was a coating or film on the inside of the cartridge that rubbed off onto the lens.After switching to one of the company's lenses, which was noticeably more "tacky" than another model lens from a same company, they observed an increase in this occurrence.
 
Manufacturer Narrative
A sample device was not returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
On initial mdr the evaluation method code of 866 was an error.It should have been g04044 on the original mdr the product was not returned.Complaint history and product history records could not be reviewed because the reporter did not provide a lot number or any identification traceable to the manufacturing documentation.A lens model was indicated but the diopter was not provided.It cannot be determined if qualified associated products were used.The product investigation could not identify a root cause for the reported complaint.This file represents the previous occurrences that they felt we due to a tech loading or folding error.The product was not returned.Not enough information was provided for further investigation.A root cause cannot be determined without physical examination of the product.Information was provided in the related file that the cartridge was only filled halfway with viscoelastic.The ifu(instructions for use)instructs to completely fill the cartridge with ovd(ophthalmic viscosurgical device) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18433420
MDR Text Key331787767
Report Number1119421-2024-00015
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CLAREON IOL.
-
-