SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Catalog Number 10697306
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2023
Event Type  malfunction  
Manufacturer Narrative
The customer provided instrument files for further investigation.Investigation is underway.Several d3 slope errors and d2 excessive drift errors were observed after the installation of the new measurement cartridge the day prior to the discrepant results.The cause of this event is unknown.
 
Event Description
The customer reported that their rp500 instrument produced discrepant sodium (na+) results when compared to another rp500 on several patient samples.Customer notes the measurement cartridge was changed due to observed d3 slope errors with sodium.There is no report of injury due to this event.
 
Manufacturer Narrative
Siemens healthcare diagnostics inc.Has confirmed an issue with rapidpoint 500 systems measurement cartridges (with lactate).This issue has the potential to affect the sodium (na+) sensor, as well as cause a question result ¿-----?¿ error flag for multiple electrolytes on patient samples and quality control.When the na+ sensor is affected, na+ slope calibration errors (d3) will be seen in the recall events log during cartridge initialization, resulting in low na+ values.The maximum sodium bias on internal testing and reported by customers comparing results to other direct method analyzers was <10 mm.To date, the impact to na+ results has been observed in less than 1% of the rapidpoint 500 systems measurement cartridges (with lactate).The question result ¿-----?¿ error flag has been observed on electrolytes.This may occur at any time over the life of the cartridge.Based on the investigation, these issues are due to an electronic noise that is observed on rapidpoint 500 systems measurement cartridges (with lactate).Measurement cartridges without lactate are not affected.Siemens has released a notification "potential for low na+ values and question result error flags" to inform customers of the issue and provide workaround options.Crr# 3002637618-11-01-2023-0004.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
UPDATE ACCORDINGLY
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18436784
MDR Text Key331810426
Report Number3002637618-2024-00108
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10697306
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberCRR# 3002637618-11-01-20