It was reported that a 30-25mm amplatzer talisman pfo occluder was selected for implant on (b)(6) 2023.The device was prepared properly per ifu.During implant it was observed that when attempting to turn the cable, the cable would not turn on the screw.The device was unable to be released.The device was removed from the patient.Outside of the patient, it was noted that there was tissue on the cable.There was no suspected tissue damage.The device was replaced with a new 30mm amplatzer multi-fenestrated septal occluder - cribriform.The patient status was stable.
|
An event of patient device interaction problem and separation problem were reported.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The reported separation problem was unable to confirm.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from field indicated that during the procedure, the cable would not turn on the screw and unable to be released the occluder.Therefore, the occluder was removed and a new occluder was used.There was tissue on the cable and there was no suspected tissue damage.Additional, a photo was provided and the occluder was attached to the cable.Based on the information received and returned device analysis, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|