It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) customer called on behalf of the bedside rn, for assistance with several error codes on console.The balloon used was an axillary inserted sensation plus 50cc that had been in for 11 days.Customer had reported that they had been dealing with condensation in the helium line for several hours and the console alarmed "error code 4" then system failure.It was reported they switched out the console.There was no patient harm reported.
|
A getinge field service engineer evaluated the unit and replaced screw kit, cardiosave preventive mainten, assy,helium reservior, pressure transducer, 30 psia, 4"cable, drive manifold assembly, pneumatic module assy, rohs, compressor, fitting,muffer to reservoir, o-ring, buna, 1 muffler,1/8 npt, muffler <100 micron, screw kit, cardiosave pneumatic assembly and passed all functional and safety tests.
|