Please note correction to g1 (manufacturing site).The reported event could be confirmed, based on the images provided by third party/user.The device inspection revealed the following: in visual inspection of the provided pictures, it was observed that the screw is broken into pieces and the breakage is near to the tip area, which could be resulted due to an application of excessive mechanical force during use.Furthermore, investigation is not possible as the device is not returned and discarded by the hospital facility.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labelling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by an application of excessive mechanical force during use.If any further information is provided, the complaint report will be updated.
|