MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number AFAPRO28 |
Device Problem
Off-Label Use (1494)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected a compromised outer vacuum.The electrical umbilical cable was reconnected and the balloon catheter was replaced but the issue persisted.The electrical umbilical cable was then replaced, but the system notice persisted.Another inflation was performed which resolved the issue. the case was completed with cryo.After the case it was reported that the balloon catheter was used after the expiration date.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the data files were returned and analyzed.One patient data file was received and recorded on the reported date of the event.The patient data file showed seven applications were performed using an afapro28 balloon catheter with lot 18846.The patient data file showed a system notice 50032 was received indicating that the safety system detected a compromised outer vacuum during applications four, five and six.The patient data file showed eleven applications were performed using a second afapro28 balloon catheter with lot 18846.The patient data file showed a system notice 50032 was received indicating that the safety system detected a compromised outer vacuum during applications nine, thirteen, fifteen and sixteen.In conclusion, the use of the device after the expiration date cannot be assessed through data analysis.The physical product is still expected to be returned for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the afapro28 balloon catheter with lot number 18846 was returned and analyzed.The balloon catheter was visually inspected and functionally tested.No anomalies were identified during external visual inspection of the balloon, shaft, and handle segments.The balloon catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for seven applications on the reported event date.The catheter was recognized and passed the electrical integrity verifications and performance test.The catheter completed the inflation, ablation, and thawing phases with no console system notices generated.No performance issues were identified.All pressure and flow were in range and temperature curve had no oscillation or overshoots.The balloon catheter was expired (exceeded its shelf life) when the procedure was performed.In conclusion, it was confirmed that the balloon catheter was expired when the procedure was performed and the balloon catheter passed the return inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|