It was reported that there was a patient involved but no injury was caused and the therapy was continuous with a constant targeted core temperature maintained.Per review of wo, it was stated that the a low flow of 1.6m/s was observed but this was isolated as a faulty gel pad.The top-level root cause of a low flow in circulation was not confirmed.All functionality tests were conducted, and all passed.It was noted that no other observations were noted apart from the suspected fault in the gel pad.
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|