Model Number 71992-01 |
Device Problem
Difficult to Insert (1316)
|
Patient Problems
Hyperglycemia (1905); Loss of consciousness (2418)
|
Event Date 01/15/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor kit was reviewed and the dhrs showed the libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
The customer reported that during the adc device application, the sensor failed to insert, as it "did not detach from sensor".As a result, the customer was unable to obtain readings therefore not being unable to monitor their glucose levels.The customer experienced a loss of consciousness and was seen at a hospital where they were hospitalized for 4 days.At the hospital, a laboratory and c-scan were performed (unspecified results) and the customer was treated with intravenous insulin (type/dose unknown) for the diagnosis of hyperglycemia.In addition, the customer was provided "lebital, lavanteted, couriga, aspirin, and asbasitin" for treatment.No further information was provided.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|
|