Model Number CD3371-40QC
Device Problems Over-Sensing (1438); Under-Sensing (1661); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that undersensing and electromagnetic interference resulting in oversensing and pacing inhibition was observed on the device.No allegation of malfunction was made against the device.Abbott technical support was contacted and confirmed the event.No intervention has been performed at this time to resolve the event.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18453853
MDR Text Key332193694
Report Number2017865-2024-00497
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberCD3371-40QC
Device Lot NumberA000094973
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-