|
Model Number 1456Q/86 |
Device Problems
Difficult to Insert (1316); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the left ventricular (lv) lead was difficult to insert into the catheter during the implant procedure.Additionally, during the slitting process, the insulation of the lead was damaged.The lead was explanted and replaced and the patient was in stable condition.
|
|
Manufacturer Narrative
|
As received, a complete lead was returned in one piece.A visual inspection revealed cuts and damage to the lead body insulation at the middle region consistent with procedural damage.The cause of the slitter has damaged the lv lead insulation is consistent with procedural damage.The reported event of difficult to push the lead forward in the sheath was not confirmed.The catheter used in the field was not returned with the lead for analysis.The catheter insertion test was performed, and the lead could be inserted inside the catheter and removed without difficulties.Furthermore, electrical testing revealed no indication of conductor fractures or internal shorts.S-curve height was measured to be within required performance specification.
|
|
Search Alerts/Recalls
|
|
|