Model Number CL-325-10-K
Device Problems Signal Artifact/Noise (1036); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted product was not returned to neuropace for analysis.
 
Event Description
During a clinic preparation, neuropace observed electrographic signal changed to a flat line on (b)(6) 2023 on lead 2.No information was available regarding patient falls, or injuries.Impedance on contact 1 on lead 2 seemed already high at the patients last follow-up on (b)(6) 2023, though signal for lead 2 looked.On (b)(6) 2023 the programming was adjusted on the lead by disabling stimulation and a change to detection.On (b)(6) 2023, a lead revision was performed, the lead was disconnected from the neurostimulator and a previously implanted lead was then connected to the rns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
5108822607
MDR Report Key18456123
MDR Text Key332735145
Report Number3004426659-2022-00071
Device Sequence Number1
Product Code PFN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCL-325-10-K
Device Catalogue Number1007933
Device Lot Number31247-1-1-1
Date Manufacturer Received11/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age25 YR
Patient SexMale
-
-