• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.H3 - other: device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device failed during the electric safety test (ground resistance test).No patient involvement has been reported.
 
Manufacturer Narrative
The following fields were updated with additional information: d9.Device available for evaluation: yes.D9.Date returned to mfg: 19-dec-2023.Investigation summary: the affected device was received for investigation.The reservoir gasket was worn out, the water tank cover was cracked, the enclosure was cracked around the quick connect, the microswitch was not working, the reflux plug/hl-390 was missing and the quick connect was corroded.After visual inspection, the reservoir was filled with water, the temp check e5581 was attached, the power cord was plugged in and the power turned on.The device failed the ground bond part of the electrical safety test, confirming the customer's indicated failure.The root cause was due to the faulty line cord, which was replaced.The water tank cover, enclosure, microswitch, reflux plug/hl-390, quick connect, reservoir gasket, o-rings, membrane switch label, and hl-390 switch label were all replaced as well.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18457999
MDR Text Key332505672
Report Number3012307300-2024-00105
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-