Model Number MDT-LEAD |
Device Problems
Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/03/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the right atrial (ra) lead exhibited high thresholds causing exit block.High and jumping impedance was also noted with oversensing and reproducible artifacts.Lead fracture was confirmed.The lead was inactivated and remains in patient. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the atrial pacing lead was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the atrial pacing lead was variable.Analysis of the device memory indicated atrial pacing capture threshold was elevated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|