This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material could not be identified nor the root cause of it.It was also discovered that there was insufficient angulation which was presumed to have been caused by excessive stress.The events can be detected/prevented in accordance with the following instructions for use: chapter 6 compatible reprocessing methods and chemical agents chapter 7 cleaning, disinfection, and sterilization procedures also, about handling the product, we confirmed the contents of the instruction manual and found the description below.From the description, the phenomenon (2) could be prevented.Warnings and cautions do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Olympus will continue to monitor field performance for this device.
|