• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA EXCHANGE SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT SPECTRA OPTIA; SPECTRA OPTIA EXCHANGE SET Back to Search Results
Catalog Number 4122201
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Itching Sensation (1943); Paresthesia (4421)
Event Date 12/14/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported to terumo bct customer support that during a therapeutic plasma exchange (tpe) procedure a patient had a citrate reaction and they had to pause the procedure.The patient experienced tingling and generalized itching and were given benadryl 50 mg ivp and calcium gluconate 2g in 100 normal saline ivpb (patient received about 75 mls on separate iv).Additionally they were given a normal saline bolus of 200 mls.The patient is in stable condition.The collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: a disposable complaint history search was performed for this lot and found no reports for similar issues on this lot.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Some of the most common reactions include fever, urticaria, hypocalcemic symptoms, pruritus, dyspnea, tachycardia, and mild hypotension.Transient hypocalcemia associated with apheresis is usually well tolerated.Symptoms often show as paresthesia (tingling) but patients may also experience unusual taste, nausea, lightheadedness, shivering, and tremors.Severe hypocalcemia may also cause muscle contractions and can progress to tetany and seizures if hypocalcemia escalates and is not corrected.According to therapeutic apheresis: a physician's handbook, adverse events occur during therapeutic procedures with a frequency of 4.8%.Symptoms of these allergic reactions may include hives, dyspnea, wheezing, burning eyes, tachycardia, hypotension, and or facial swelling and flushing.Mild reactions can be treated with diphenhydramine administered through an iv.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.Root cause: a root cause assessment was performed for the reported citrate reactions.These reactions occur due to decreased ionized calcium in circulation as a result of exogenous citrate administered during the apheresis procedure and are influenced by patient physiology, the rate of ac infusion, and/or the length of the procedure.These symptoms may be treated with oral or intravenous calcium supplements or by adjusting the ac infusion rate.A root cause assessment was performed for the allergic reactions.Based on the available information a definitive root cause could not be determined but it is likely due to one or a combination of the possible causes listed below: * hypersensitivity to the ethylene oxide used to sterilize the disposable set.* hypersensitivity to the components inside the disposable set.* allergic reactions to the replacement fluids.* anaphylactic reactions have been reported in iga-deficient patients with anti-iga antibodies and patients with haptoglobin deficiency, most reactions are idiosyncratic and not associated with a specific serum protein deficiency, polymorphism, or identifiable cause.* immediate hypersensitivity reaction (production of ige antibodies in response to allergens (plasma proteins in fresh frozen plasma).* passive sensitization (passive transfer of donor allergens or donor antibodies (e.G.Allergen-specific ige); transient allergic sensitivity to allergens.*patient¿s underlying disease state or sensitivity to the apheresis procedure.
 
Event Description
The customer reported to terumo bct customer support that during a therapeutic plasma exchange (tpe) procedure a patient had a citrate reaction and they had to pause the procedure.The patient experienced tingling and generalized itching and were given benadryl 50 mg ivp and calcium gluconate 2g in 100 normal saline ivpb (patient received about 75 mls on separate iv).Additionally they were given a normal saline bolus of 200 mls.The patient is in stable condition.The collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Type of Device
SPECTRA OPTIA EXCHANGE SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key18466472
MDR Text Key332331698
Report Number1722028-2024-00011
Device Sequence Number1
Product Code LKN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number4122201
Device Lot Number2308236241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
Patient SexMale
Patient Weight88 KG
-
-