Model Number SCD7A39
Device Problems Difficult or Delayed Activation (2577); Detachment of Device or Device Component (2907); Difficult to Open or Close (2921)
Patient Problems Tissue Breakdown (2681); Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  Injury  
Manufacturer Narrative
D10 concomitant product: scg7aa, sonicision 7 generator a scg7aa (serial#: (b)(6)) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic colectomy, when clamping the tissue, the jaw of the device did not clamp properly, moreover there was an uneven tissue cut, sometimes the device cuts, sometimes it did not.There was no problem with the hemostasis.There was no patient injury.A photo was received showing that the active waveguide of the device had chipped.
 
Event Description
According to the reporter, during a laparoscopic colectomy, when clamping the tissue, the jaw of the device did not clamp properly, moreover there was an uneven tissue cut, sometimes the device cuts, sometimes it did not.There was no problem with the hemostasis.Replaced with another device and it worked fine.A photo was received showing that the active waveguide of the device had was dented.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that when clamping the tissue, the jaw of the device did not clamp properly, moreover there was an uneven tissue cut, sometimes the device cuts, sometimes it did not.The reported issue was confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: contact between the active blade and other metal objects (hemostats, clips, staples, retractors, etc.) may result in product damage, such as a broken blade.Pieces of a broken blade may fall into the surgical cavity causing unintended tissue damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
(fdp/ime/imf: serious injury/ illness/ impairment, foreign body in patient) new information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic colectomy, when clamping the tissue, the jaw of the device did not clamp properly, moreover there was an uneven tissue cut, sometimes the device cuts, sometimes it did not.There was no problem with the hemostasis.Replaced with another device and it worked fine.Broken piece was not retrieved from patient body.A photo was received showing that the active waveguide of the device had chipped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICISION 7
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18467552
MDR Text Key332766426
Report Number1717344-2024-00058
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10884521794795
UDI-Public10884521794795
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K212301
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSCD7A39
Device Catalogue NumberSCD7A39
Device Lot Number30190027X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE NOTE ON H10
Patient Outcome(s) Other;
-
-