• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE SOFT; FDI, SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE SOFT; FDI, SNARE, FLEXIBLE Back to Search Results
Catalog Number SASMH-1-S
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Section g: pma/510(k): k173673.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use (ifu) states: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." instructions for use states: "fully retract and extend snare to confirm smooth operation of device." instructions for use states: "inspect active cord.Cord must be free of kinks, bends, breaks, and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an unknown endoscopic procedure, the physician used multiple cook acusnare polypectomy snare.It was reported that the snare handle is making the snares very rigid and difficult to close and not cutting.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSNARE POLYPECTOMY SNARE SOFT
Type of Device
FDI, SNARE, FLEXIBLE
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key18470395
MDR Text Key332988926
Report Number1037905-2024-00018
Device Sequence Number1
Product Code FDI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSASMH-1-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN ENDOSCOPE MAKE AND MODEL.
-
-