Model Number LTF-S190-5 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned and evaluated, and there was found to be damage on charged coupled device unit.Additional findings include the following: breakage of the circuit board, adhesive on bending section cover has a chip, deformation or breakage due to physical stress (handling problem), video connector has a crack.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, the endoeye flex deflectable videoscope image was not transferred.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three good faith effort (gfe) attempts were made to retrieve additional information.Customer did not respond.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that phenomenon 1 "no image" and phenomenon 2 "id function malfunction" were caused by damage to the image sensor unit (e.G., disconnection) or failure of mounted components (integrated circuit (ic) chip, capacitor, etc.) on the electrical board due to stress from use, external factors, or handling.The event can be detected/prevented by following the instructions for use which state: the inspection method for the suggested event is described as follows in the operation manual "chapter 3 preparation and inspection 3.8 inspection of the endoscopic system¿.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|