Model Number LTF-S190-5
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated, and there was found to be damage on charged coupled device unit.Additional findings include the following: breakage of the circuit board, adhesive on bending section cover has a chip, deformation or breakage due to physical stress (handling problem), video connector has a crack.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the endoeye flex deflectable videoscope image was not transferred.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three good faith effort (gfe) attempts were made to retrieve additional information.Customer did not respond.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that phenomenon 1 "no image" and phenomenon 2 "id function malfunction" were caused by damage to the image sensor unit (e.G., disconnection) or failure of mounted components (integrated circuit (ic) chip, capacitor, etc.) on the electrical board due to stress from use, external factors, or handling.The event can be detected/prevented by following the instructions for use which state: the inspection method for the suggested event is described as follows in the operation manual "chapter 3 preparation and inspection 3.8 inspection of the endoscopic system¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18471780
MDR Text Key332400035
Report Number9610595-2024-00476
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434648
UDI-Public04953170434648
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K201832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Was the Report Sent to FDA? No
Date Manufacturer Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-