ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD2357-40Q
Device Problems Failure to Capture (1081); High impedance (1291); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  Injury  
Event Description
It was reported that the patient presented in clinic for follow up.Upon interrogation it was noted that the implantable cardioverter defibrillator (icd) was exhibiting high defibrillation impedance, high pacing impedance, decreased p-wave amplitude sensing and high capture thresholds that resulted in failure to capture.The icd was explanted and new icd was implanted.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of impedance and sensing anomaly was confirmed.The implantable cardioverter defibrillator (icd) was returned for analysis.Interrogation of the device revealed the device was above the elective replacement indicator (eri) when received.Visual inspection of the device did not reveal any anomaly that could contribute to the reported event.During testing, the failure was lost and could not be reproduced.The cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18472053
MDR Text Key332373707
Report Number2017865-2024-00683
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Model NumberCD2357-40Q
Device Lot NumberS000079459
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA.; TENDRIL STS.
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient SexMale
Patient Weight64 KG
-
-