Catalog Number UNK REGENT HEART VALVE |
Device Problems
Fracture (1260); Detachment of Device or Device Component (2907)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/15/2023 |
Event Type
Injury
|
Event Description
|
It was reported that on (b)(6) 2023, an unknown regent valve was chosen for implantation.After implant, when the valve was rotated it broke, resulting in one of the leaflets falling into the ventricle.The leaflet was retrieved and a replacement mechanical heart valve was implanted.
|
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Manufacturer Narrative
|
An event of leaflet detachment and fracture was reported.The possible causes of the reported event include an external force applied to the valve leaflets that may cause structural damage, the use of a hard or rigid instruments to test leaflet mobility and oversizing of the valve.Field indicated that the valve was rotated during implantation which could have contributed to the reported event.Based on available information and without the device to analyze, the root cause of the reported event could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
|