Model Number PV2015L20-A
Device Problem Material Fragmentation (1261)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/12/2023
Event Type  Injury  
Event Description
During insertion of a femoral picco arterial line using the seldinger technique it was noticed that the guide-wire was faulty.On removal the wire had separated and a 10cm section of fine wire had broken off-likely inside the patient.
 
Manufacturer Narrative
Manufacturer reference#: (b)(4).
 
Manufacturer Narrative
It has been reported that the guidewire had been broken and a 10cm section of the guidewire remained inside the patient.A provide picture showed a uncoiled guidewire.After several failed attempts to send the device for evaluation, a retained sample was investigated visually.The retained sample of the same lot (722789) was visually analyzed.No deviation could be detected.A dhr review did not reveal any non conformity or deviation from specification which could have contributed to the event.There is no indication for a systematic root cause as a deficiency of design, production or material considering the very low complaint rate (b)(4) , considering all types of picco catheters).One similar complaint has been received, reporting a guidewire separation and remaining parts in the patient.Unfortunately, the product in question was not available for analysis.Therefore, it is not possible to determine whether the device has failed to meet its specifications when the event occurred.Considering worst case scenario, a relationship between the device and the complaint is therefore considered as likely.Information indicates that upon the complaint occurrence the device was involved and used on a patient for advanced hemodynamic monitoring.Upon the event occurrence the device was involved and catheter insertion procedure on a patient for advanced hemodynamic monitoring was ongoing.The accessory guidewire is supplied in the product picco catheter.The procedure to insert a catheter (¿seldinger technique¿) is well known for healthcare professionals.The issue is monitored on a regularly basis in order to detect early trends.As there is no trend for this kind of issue no additional actions will be taken.With the information stated above the complaint will be closed.Correction: of udi from (b)(4), h3 other text : device requested but not returned.
 
Event Description
Manufacturer reference #: 958746.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICCO CATHETER
Type of Device
PROBE, THERMODILUTION
Manufacturer (Section D)
PULSION MEDICAL SYSTEMS SE
hans-riedl-strasse 17
feldkirchen 85622
GM  85622
Manufacturer (Section G)
PULSION MEDICAL SYSTEMS SE
hans-riedl-strasse 17
feldkirchen 85622
GM   85622
Manufacturer Contact
diana kitschke
hans-riedl-strasse 17
feldkirchen 85622
GM   85622
0894599140
MDR Report Key18481963
MDR Text Key332491191
Report Number3003263092-2024-00001
Device Sequence Number1
Product Code KRB
UDI-Device Identifier14250094500938
UDI-Public(01)14250094500938
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171620
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPV2015L20-A
Device Catalogue Number6885049
Device Lot Number722789
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/07/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-