Siemens has completed the investigation.The rp500e instrument including the co-oximetry subunit, was found to be performing as intended.There were no system or sensor specific errors in and around the time of the patient sample in question was measured.The automatic quality control (aqc) data for the total hemoglobin (thb) parameter was stable for the duration of measurements and within reportable range.For the patient sample in question, the sensor raw responses were reviewed, and no anomalous behavior was noted.The sample passed all internal sample and end-point detection quality checks.Thb assays are also known to be highly sensitive to sample collection and handling techniques.Preanalytical factors like hemolysis and insufficient mixing, time to sampling, etc.Are known causes for imprecise thb results.The following facts were noted that cannot confirm the alleged discrepancy.The thb result from the secondary instrument was run 90 minutes prior to the alleged patient sample in question.The whole panel of parameters was not available to review to observe if the samples were comparable.While it is reported that the discrepancy was confirmed with repeat samples, there were no repeat samples run on the escalated instrument nor the data from the repeat sample was shared to review.The cause of this event is unknown.No device problem found.
|