BOSTON SCIENTIFIC CORPORATION RESOLUTION 360 CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT Back to Search Results
Model Number M00521230
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2023, for the treatment of closing a bleed.During the procedure, it was reported the technician couldn't deploy the third clip from the catheter.Another technician tried but was also unable to release the clip.The first technician opened his hand up so wide and hard that the handle's plastic started to crack.The procedure was completed with another resolution 360 clip device.There were no patient complications have been reported as a result of this event.No further information has been obtained despite good-faith efforts.
 
Manufacturer Narrative
Block h6: imdrf device code a15 captures the reportable event of unable to deploy the clip off the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION 360 CLIP
Type of Device
HEMOSTATIC METAL CLIP FOR THE GI TRACT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18485807
MDR Text Key333097005
Report Number3005099803-2023-07128
Device Sequence Number1
Product Code PKL
UDI-Device Identifier08714729875642
UDI-Public08714729875642
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00521230
Device Catalogue Number54774
Device Lot Number0032675813
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-