H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the rem alarm is lit even when another return electrodes is used, so it was determined to be a problem with the main unit.Functionally, the rem indicator was green even when the return electrodes was not connected due to a defective rf board.It was reported that the unit had rem alarm abnormality.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
According to the reporter, prior to use, the unit had rem alarm abnormality.The rem signal does not remain green when it should not.The rem alarm is lit even when another return electrodes is used, so it was determined to be a problem with the main unit.Return electrodes is a product of another company, so it is unknown.Monopolar was not used (not possible), only ligasure was used.There was no patient involvement.Medtronic's initial evaluation of the incident device found the rem indicator was green even when the return electrodes was not connected due to a defective rf board.
|