Catalog Number 5C4482
Device Problems Loose or Intermittent Connection (1371); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter facility name:.E1: initial reporter city:.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a connection issue between the patient connector of the minicap extended life pd transfer set and the titanium adapter.This was observed during use of the devices for peritoneal dialysis therapy.The transfer set was replaced; however, the titanium adapter remained in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
B5: there was a loose connection between the transfer set and adapter.H10: the actual device was not available; however two (2) photographs of the sample were provided for evaluation.A review of the returned photographs did not show evidence of the reported problem.Due to the nature of the sample, no further evaluation could be performed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18487568
MDR Text Key333035467
Report Number1416980-2023-07052
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4482
Device Lot NumberH23C23058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TITANIUM ADAPTER
-
-