Model Number 6662 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problems
Failure of Implant (1924); Inadequate Pain Relief (2388); Movement Disorder (4412)
|
Event Date 12/20/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on (b)(6) 2017.
|
|
Event Description
|
It was reported that following a procedure the ipg did not communicate with external devices.Reportedly, the ipg was not set into surgery mode prior to the procedure and electrosurgical device was used during the procedure.Next course of action is pending.
|
|
Manufacturer Narrative
|
Corrected data: e1: reporter phone number and h6 health effect - clinical code the results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information received indicates that the procedure after which the ipg got inoperable was not related to the abbott system.Surgical intervention may take place at a later date to address the issue.
|
|