Catalog Number 309657 |
Device Problem
Missing Information (4053)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd luer-lok syringe had a scale marking issue.The following was provided by the initial reporter: "removed 3ml syringe from package and found that there were not numbers or dashes to determine how many mls were placed in syringe." type of product 3 ml sterile syringe product ref # (b)(4).Product lot # 3242479.
|
|
Manufacturer Narrative
|
Pr 9448212¿ follow up mdr for device evaluation one sample and one photo of a 3 ml luer-lok syringe were received by bd.A quality engineer was able to review the sample and photo from lot number 3242479 regarding material number 309657.The sample was received with the package with all applicable product information on the top web and syringe missing all scale markings.The image shows a loose syringe with all the scale markings missing.The condition observed is non-conforming per product specification.Potential root cause for the missing print defect is associated with the marking process.A device history record review was completed for provided lot number 3242479 showing no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10 manufacture narrative.
|
|
Event Description
|
No additional information.
|
|
Search Alerts/Recalls
|