Catalog Number 309657
Device Problem Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that the bd luer-lok syringe had a scale marking issue.The following was provided by the initial reporter: "removed 3ml syringe from package and found that there were not numbers or dashes to determine how many mls were placed in syringe." type of product 3 ml sterile syringe product ref # (b)(4).Product lot # 3242479.
 
Manufacturer Narrative
Pr 9448212¿ follow up mdr for device evaluation one sample and one photo of a 3 ml luer-lok syringe were received by bd.A quality engineer was able to review the sample and photo from lot number 3242479 regarding material number 309657.The sample was received with the package with all applicable product information on the top web and syringe missing all scale markings.The image shows a loose syringe with all the scale markings missing.The condition observed is non-conforming per product specification.Potential root cause for the missing print defect is associated with the marking process.A device history record review was completed for provided lot number 3242479 showing no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10 manufacture narrative.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18489752
MDR Text Key333009595
Report Number1213809-2023-01530
Device Sequence Number1
Product Code FMF
UDI-Device Identifier30382903096573
UDI-Public(01)30382903096573
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110771
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number309657
Device Lot Number3242479
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-