Model Number RNS 320 |
Device Problems
Failure to Interrogate (1332); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2023 |
Event Type
malfunction
|
Event Description
|
The patient underwent an mri on (b)(6) 2023, the rns was placed in mri enabled mode prior to the mri.The treating clinician attempted to place the device back to normal operating mode, but was unable to communicate with the device.Troubleshooting was performed along with support from neuropace field clinical engineering but it was determined that the device was no longer functional and needed to be replaced.The device was replaced without complication on (b)(6) 2023.
|
|
Manufacturer Narrative
|
(b)(4) the explanted device was returned to neuropace and is undergoing analysis.
|
|
Event Description
|
Original report: the patient underwent an mri on (b)(6) 2023, the rns was placed in mri enabled mode prior to the mri.The treating clinician attempted to place the device back to normal operating mode, but was unable to communicate with the device.Troubleshooting was performed along with support from neuropace field clinical engineering but it was determined that the device was no longer functional and needed to be replaced.The device was replaced without complication on (b)(6) 2023.
|
|
Manufacturer Narrative
|
(b)(4).This device failed due to exposure to an mri procedure.Such a procedure is known to expose the device to strong fields that induce a voltage on the patient leads.The cinco 1e asic internal power supplies can be susceptible to failure under these conditions.Though there is no evidence of physical damage to the device this failed state is unrecoverable without gaining physical access to the circuit.
|
|
Search Alerts/Recalls
|