|
Catalog Number EMAX2PLUS |
Device Problems
Device-Device Incompatibility (2919); Fail-Safe Problem (2936)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2023 |
Event Type
malfunction
|
Event Description
|
It was reported from china that during service and evaluation, it was determined that the motor device cutter could not be secured and ran in the locked position.It was further determined that the device failed pretest for cutter lock and safety.It was noted in the service order that the device had a lock issue.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component wear.Udi:(b)(4).
|
|
Search Alerts/Recalls
|
|
|