It was reported that a mitraclip procedure was performed to treat a functional mitral regurgitation (mr) with grade of 4+.An ntw clip was implanted with no reported issue, reducing mr to grade 1-2.During removal of the clip delivery system (cds), and the physician was securing the clip introducer while pulling the cds from the steerable guide catheter (sgc), the flush port of the clip introducer broke off.The removal steps were performed per instruction for use (ifu).No additional force was applied during removal.There was no harm to the patient.There was no clinically significant delay in the procedure and no adverse patient effects.
|
All available information was investigated and the reported clip introducer (ci) flush port was confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported broken ci flush port.There is no indication of a product issue with respect to manufacture, design, or labeling.
|