MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY 4085 SURGICAL TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY 4085 SURGICAL TABLE Back to Search Results
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the table and found that the table height sensor was not reading properly resulting in the reported event.The 4085 surgical table was installed in 2017 making it approximately 7 years old and is not under steris service agreement for maintenance activities.The user facility is responsible for all maintenance activities.The user facility elected to perform the repairs themselves and ordered a replacement table height sensor.The table has been removed from service until the repairs can be completed.No additional issues have been reported.
 
Event Description
The user facility reported that their 4085 surgical table lowered in height by approximately two inches during a patient procedure.The procedure was completed successfully.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4085 SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key18495888
MDR Text Key333198754
Report Number1043572-2024-00003
Device Sequence Number1
Product Code FQO
UDI-Device Identifier00724995154783
UDI-Public00724995154783
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-