Model Number BAR-A8-8MM
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  Injury  
Event Description
The surgeon was concerned about proximity to the nerve root and did not align the occlusion component with the defect.Significant occlusion component buckling encountered during implantation.Implant was removed prior to being fully implanted and countersunk.Removal was done successfully with no issues.Image in sc shows a fractured nitinol guide from one end.Removal tool was utilized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRICAID ACD
Type of Device
BARRICAID
Manufacturer (Section D)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer (Section G)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer Contact
rachael nelsen
30 commerce way
woburn, MA 01801
MDR Report Key18497813
MDR Text Key332700382
Report Number3006232063-2024-00002
Device Sequence Number1
Product Code QES
UDI-Device IdentifierM906BARA88MM0
UDI-PublicM906BARA88MM0
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBAR-A8-8MM
Device Catalogue Number2730815-A8
Device Lot Number06152302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-