• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Manufacturer Narrative
The distributor performed a checkout of the equipment and confirmed the reported complaint.The customer declined service.No repair information available.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.H3 other text : the distributor performed a checkout of the equipment and confirmed the reported complaint.The customer declined service.No repair information available.Block a: no report of patient involvement.
 
Event Description
It was reported that there was a malfunction resulting in a loss of mechanical ventilation.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key18498991
MDR Text Key333138593
Report Number9710602-2024-00069
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/15/2023
Date Device Manufactured05/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-