MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP® SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP® SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0301
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) grade 4+.A xtr mitraclip was advanced to the valve.During manuevering, the clip became entangled in the chorae.The clip could not be removed from the chordae.Troubleshooting was attempted but was unsuccessful.There was a transient pericardial effusion due to the chordal entanglement.The patient was transferred to open heart surgery to retrieve the mitraclip and replace the mitral valve.When the mitraclip system was removed, the tip of the steerable guide catheter (sgc) was noted to be damaged.
 
Manufacturer Narrative
In this case, the returned device analysis was unable to confirm the reported deformation due to compressive stress associated with deformed soft tip.The distal shaft however, was noted to be deformed (kinked) close to the distal marker band by the tip ring.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and a cause for the reported deformed (kinked) distal end of the sgc cannot be confirmed.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP® SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18499287
MDR Text Key332731311
Report Number2135147-2024-00212
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSGC0301
Device Lot Number30619R1035
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
-
-