It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) grade 4+.A xtr mitraclip was advanced to the valve.During manuevering, the clip became entangled in the chorae.The clip could not be removed from the chordae.Troubleshooting was attempted but was unsuccessful.There was a transient pericardial effusion due to the chordal entanglement.The patient was transferred to open heart surgery to retrieve the mitraclip and replace the mitral valve.When the mitraclip system was removed, the tip of the steerable guide catheter (sgc) was noted to be damaged.
|
In this case, the returned device analysis was unable to confirm the reported deformation due to compressive stress associated with deformed soft tip.The distal shaft however, was noted to be deformed (kinked) close to the distal marker band by the tip ring.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and a cause for the reported deformed (kinked) distal end of the sgc cannot be confirmed.There is no indication of a product issue with respect to manufacture, design or labeling.
|