ABBOTT MEDICAL MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR); CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0701
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 12/21/2023
Event Type  Injury  
Event Description
It was reported that a mitraclip procedure was performed to treat a functional mitral regurgitation (mr) with grade of 4+.When the steerable guide catheter (sgc) was in the left atrium, and the physician was trying to locate where it was, tissue flopping on the wall of the atrium was observed.The cause of the flopping tissue could not be determined.No further action was taken with the tissue.An xtw clip was advanced to valve.When maneuvering towards the valve, they saw what they thought was a thrombus on tip of the clip.Further inspection performed, but the object was gone and there was no thrombus observed.The clip was implanted successfully, reducing mr to grade 2+.Patient was fully heparinized with act over 250 seconds.There was no clinically significant delay in the procedure and no adverse patient sequelae.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Based on the information reviewed, a cause for the unspecified tissue injury (tissue flopping on the wall of the atrium) cannot be determined.Tissue damage is listed in the instructions for use (ifu) as a known possible complication associated with mitraclip procedures.Serious injury/illness/impairment was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18502701
MDR Text Key332776593
Report Number2135147-2024-00224
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberSGC0701
Device Lot Number30914R1116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
Patient SexFemale
Patient Weight51 KG
-
-