Updated fields: b4, d9, g3, g6, h2, h3, h4, h6, h10.A getinge field service engineer (fse) was dispatched to investigate the issue.After the inspection, the fse found that the iabp failed the helium leak check by dropping more than 20psi in 5 minutes.The fse also found that when the iabp console was removed from the hospital cart, more than 3 psi was lost in 30 minutes from the internal tank.The fse replaced the helium reservoir (d997-00-0565).After installing the part, the helium leak was no longer present.The fse then performed a pm, full calibration, and tested the unit.The product passed all functional/safety testing and was returned to the customer.
|