Dhr was reviewed, and the pump passed all previous tests.There are no previous complaints on this device.The pump was received on 1/24/2024 and tested on (b)(6) 2024.The results are below: the review of the event log confirms that the pump experienced a system error alarm during an infusion.This is seen by the line that indicates alarm code:02 sub code 44.The subcode means motor open, motor delay issue.The full event log is attached to the record.See below: event 70: log_event_alarm, time: 22:12:49, alarm code: 02, sub code: 24.The reported issue is confirmed.The pump does not meet passing criteria.
|