Model Number M00510890
Device Problems Difficult to Open or Close (2921); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, upon pulling the handle, the basket did not fully expand.The procedure was completed with another trapezoid rx.There were no patient complications reported as a result of this event.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the side car rx was pushed back.Please see block h10 for full investigation details.
 
Manufacturer Narrative
Block e1: (b)(6).Block h6: imdrf device problem code a0406 captures the reportable investigation finding of side car rx push back.Block h10: the returned trapezoid rx lithotripter basket was received for analysis, and a visual inspection observed that the side car rx was pushed back and the sheath was buckled.The device was filled with remnants of use.A dimensional test confirmed that the side car rx was pushed back approximately 2.5 mm, which is out of specification.Additionally, a functional test was performed by attempting to open the basket, and it was found the basket was only able to open properly after cleaning out the remnants of use inside the device.No other issues were noted.The reported event was confirmed.The results of the analysis performed showed that the basket wasn't opening at all initially when using the handle.After the remnants of use were cleaned, the basket was able to open without any major difficulties.The sheath was also found buckled, which could have happened as a result of the force applied on the handle by trying to open the basket.The side car rx may also be affected when excessive force is applied to the handle.In this case, since the basket wasn't being opened and was stuck, this pressure affected the side car rx causing it to push back.Therefore, the most probable root cause for the reported issue and the issues found during analysis is adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18503303
MDR Text Key333157347
Report Number3005099803-2023-07066
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0031845038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-