• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSAR, INC LENSAR LASER SYSTEM-FS 3D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSAR, INC LENSAR LASER SYSTEM-FS 3D Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Capsular Bag Tear (2639)
Event Date 12/12/2023
Event Type  Injury  
Manufacturer Narrative
Procedure id (b)(6) was reviewed.It does appear they docked a few times causing a film on the bottom of the surgical window.The capsular button appeared to have been removed easily with no adhesions.Surgical footage indicates the tear happened during hydrodissection.Findings were communicated to the doctor.
 
Event Description
P1008 - posterior radial tear -issue: on (b)(6) 2023, (lls2223) reported to cas that dr.Experienced a posterior tear on his last case of the day on (b)(6) 2023 on procedure #1966.Per dr., this has been an ongoing problem despite various laser setting changes and support visits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSAR LASER SYSTEM-FS 3D
Type of Device
LENSAR LASER SYSTEM-FS 3D
Manufacturer (Section D)
LENSAR, INC
2800 discovery drive ste.100
orlando FL 32826
Manufacturer (Section G)
LENSAR, INC
2800 discovery drive ste. 100
orlando FL 32826
MDR Report Key18503950
MDR Text Key332879022
Report Number3009026057-2023-60130
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182795
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number70-00005-001
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-