This report has been identified as b.Braun internal report number (b)(4).Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.1.Visual inspection.The complaint sample was not returned.The condition of complaint sample was confirmed from the photo attached in (b)(4).It was observed that the cannula of complaint sample was torn at a position approximately 70 mm from the base of hub.2.Functional test.Result: no abnormality.Bending stiffness test was conducted on the retention sample of cannula pipe in accordance with the testing and measuring method no.: im-2011.The test results met our standards.3.Others.[visual inspection of retention sample].There were no abnormalities such as bent cannula on the appearance of the retention samples (n=5pcs).4.Device history record.Result: no abnormality.Any abnormality was not observed on the manufacturing records of pencan 27gx4 w.Intro.-eu/ap (batch no.22n15h8b05).5.Decision and justification.This complaint is considered as not confirmed.The complaint sample had been used for a procedure, and there were deformations in addition to the broken part.It is presumed that the cannula of the complaint sample was bent and torn during the procedure.
|