The mayfield skull clamp (a3059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the reported complaint was confirmed from the evaluation: there was rotational movement and a residue buildup present in the swivel lock assembly.To resolve all issues, the internal parts to adjust the unit to manufacturer specifications were replaced and the swivel lock assembly was cleaned.After completing the reassembly, a function test was successfully performed.Root cause - the complaint is confirmed via inspection of the unit.The unit required replacement of worn internal parts and cleaning due to routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can contribute to movement or slippage of the patient¿s head.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|