Model Number 471172-17 |
Device Problems
Thermal Decomposition of Device (1071); Material Integrity Problem (2978)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/10/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have thermal damage on the bipolar yaw pulley.Additionally, the instrument was found to have the conductor wire insulation damaged.There was material missing.Material missing measures approximately 029" x.085".As a result, the conductor wires were exposed.The instrument passed an electrical continuity test.
|
|
Event Description
|
It was reported that after a da vinci-assisted simple prostatectomy surgical procedure, a fault at the tip was found when the maryland bipolar forceps instrument was inspected.The tip appeared to be grooved or cracked.The procedure was completed.No known impact or patient consequence was reported.
|
|
Manufacturer Narrative
|
Intuitive surgical inc.(isi) followed up with the initial reporter and obtained the following additional information: it is not possible to say that the missing material/damage was outside the procedure.However, the damage was only identified in the cme during the washing process.There were no reports of fragments falling from the device while used within the patient.All patients undergoing robotic surgery are monitored by the hospital's sccih.If there are fragments, the surgeon in charge will deal with them.The patient has not returned to the hospital due to experiencing any post-surgical complications related to retention of foreign material.Thermal damage was not observed.The damaged was verified after the event in the cme sector, and it is not possible to verify if arcing was observed during the procedure.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|
Search Alerts/Recalls
|